Analyseur de mesure fixe
SIGNAL S4 SOLAR : Analyseur de COV FID chauffé
Caractéristiques
Measurement method: Flame ionisation detector
Units of measurement: PPM or mg (user-selectable)
Measurement ranges: Range A: 0–1000 ppm.
User-adjustable, e.g. 0–1 ppm, 0–5 ppm,
0–10 ppm, 0–50 ppm, 0–100 ppm, 0–500 ppm, 0–1000 ppm.
Resolution: 0.01 ppm
Range B: 0–10,000 ppm.
User-adjustable, e.g. 0–10 ppm,
0–50 ppm, 0–100 ppm, 0–500 ppm, 0–1000 ppm,
0–5000 ppm. 0–10,000 ppm.
Resolution: 0.1 ppm
Range C: 0–100,000 ppm.
User-adjustable, with a resolution of 1 ppm
Response time: THC <1.5 sec
CH₄ and NMHC <2.5 sec
Repeatability: <1% FSD
Oxygen effect: <2% of reading at 21% O₂ (H₂He)
Linearity: ± 0.5% FSD or 2% of the point
EN14181 - dc rel: <0.5
R2: >0.99
Drift: ± 0.2 ppm or 2% of the range per week, whichever is higher
Noise: ±0.1 ppm or 1% of the range, whichever is higher
Temperature effect on
Zero:
<0.15% per °C
Temperature effect on
Span:
<0.3% per °C
Sample inlet pressure: With internal sampling pump:
-0.6 to +0.4 bar
Without internal sampling pump:
+0.2 to +0.5 bar
Accuracy: <0.2% FSD
Precision: <1%
Detection limit: 0.05 mgC/m³
Bypass flow sensitivity: Less than 2% of 1 to 3 l/min
Sample filter: Removable 0.4-micron PTFE
7-micron stainless steel filter
Non-removable stainless steel for CFID
Display: Blank or detachable screen
Sample conditions: 0–200°C (heated version)
0–80°C, non-condensing for CFID
Fuel consumption: Single detector:
35 ml/min H₂ or 180 ml/min H₂He
Dual detectors:
70 ml/min H2 or 360 ml/min H2He
Air supply: Single detector:
>1.1 l/min
Dual detector:
>1.6 l/min
Operating conditions: 5–40°C ambient temperature
Outputs: 0–10 Vdc
RS232
Ethernet
TCP/IP
4-20 mA option
Power requirements: 100 to 250 VAC
24 VDC optional
600 W max
Remote control: AK protocol via RS232 port,
Ethernet
Supplied with S4i operating software.
Dimensions and weight: 19" (W) x 133.5 (H) x 530 mm (D)
Approx. 30 kg
Points forts
Flexible:
Fixed and portable versions
‘Hot’ and ‘cold’ versions
EN14181 QAL1 (MCERTS) applied for
Easy to use:
Fully automatic operation
Wireless tablet
Software suite for use via Ethernet or RS232
Accurate:
Precision monobloc FID
PPM trace measurement standard
High range, % available
GASES:
Volatile organic compounds (VOCs)
Total hydrocarbons (THC)
Non-methane hydrocarbons (NMHC)
Methane (CH4)
Applications:
CEMS
Research
Compliance
Gas purity
Automotive
Air quality
Processes
Combustion
A wide range of user-configurable alarms are available for conditions such as:
1. Concentration limit (user-defined)
2. Sample flow rate (outside limits)
3. Pump failure
4. Heater failure
5. Voltage outside limits
6. Thermocouple failure
7. EHT out of limits
8. Configuration error
9. Incorrectly set options
10. Incorrect calculations (no calibration)
Detector
Temperature
Ranges
Fixed cold FID (CFID-THC) 120°C for a 0-1 ppm detector up to 0-1000 ppm
Fixed cold FID (HFID-THC) 200°C, for a detector 0-10 up to 0-10,000 or 0-100 up to 0-100,000
Fixed hot methane and non-methane FID Dual detector at 200°C 0-10 up to 0-10,000 or 0-100 up to 0-100,000 Cutoff efficiency 98%.
Fourni avec
Flame ionisation detector (FID) analysers for gas purity, air separation plants, engine emissions, combustion studies and monitoring of VOC reduction in treatment plants.