Analyseur détecteurs portables
SIGNAL S4 SOLAR XPLORE : Analyseur FID de COV portable et chauffé
Caractéristiques
Measurement method: Flame ionisation detector
Units of measurement: PPM or mg (user-selectable)
Measurement ranges: Range A: 0–1000 ppm.
User-adjustable, e.g. 0–1 ppm, 0–5 ppm,
0–10 ppm, 0–50 ppm, 0–100 ppm, 0–500 ppm, 0–1000 ppm.
Resolution: 0.01 ppm
Range B: 0–10,000 ppm.
User-adjustable, e.g. 0–10 ppm,
0–50 ppm, 0–100 ppm, 0–500 ppm, 0–1000 ppm,
0–5000 ppm. 0–10,000 ppm.
Resolution: 0.1 ppm
Range C: 0–100,000 ppm.
User-adjustable, with a resolution of 1 ppm
Response time: THC <1.5 sec
CH₄ and NMHC <2.5 seconds
Repeatability: <1% FSD
Oxygen effect: <2% of the reading from
0% to 21% O₂ (H₂He)
Linearity: ± 0.5% FSD or 2% of the point
EN14181 - dc rel: <0.5
R²: >0.99
Drift: ± 0.2 ppm or 2% of the range per week, whichever is higher.
Noise: +/-0.1 ppm or 1% of the range, whichever is greater.
Temperature effect on zero: <0.15% per °C
Temperature effect on span: <0.3% per °C
Sample inlet pressure: With internal sampling pump:
-0.6 to +0.4 bar
Without internal sampling pump:
+0.2 to +0.5 bar
Accuracy: <0.2% FSD
Precision: <1%
Detection limit: 0.05 mgC/m³
Bypass flow sensitivity: Less than 2% of 1 to 3 l/min
Sample filter: Removable 0.4-micron PTFE
7-micron stainless steel filter
Non-removable stainless-steel filter for CFID
Display: Fixed or detachable screen
Sample conditions: 0–200°C (heated version)
0–80°C non-condensing for CFID
Gas flow rate: Single detector:
35 ml/min H₂ or 180 ml/min H₂He
Dual detectors:
70 ml/min H₂ or 360 ml/min H₂He
Air supply: Single detector:
>1.1 L/min
Dual detector:
>1.6 L/min
Operating conditions: 5–40°C ambient temperature
Outputs: 0–10 Vdc
RS232
Ethernet
TCP/IP
4–20 mA (optional)
Power requirements: 100–250 Vac
24 Vdc (optional)
600W max
Remote control: AK protocol
Ethernet
Supplied with S4i operating software.
Dimensions and weight: 19" (W) x 133.5 (H) x 530 mm (D)
Approx. 22 kg
Protection rating: IP 55
Points forts
Flexible:
Single or dual detectors
Integrated zero air generator
Software suite for use over Ethernet
Integrated heated line controller
Easy to use:
Robust and portable
Wireless interface for tablet
Typical warm-up time of 30 minutes
IP55
Accurate:
Precision monobloc heated FID.
Simultaneous monitoring of THC, NMHC and methane.
Fast response time.
Complies with standard EN12619.
Fourni avec
Flame ionisation detector (FID) for the discontinuous measurement of the mass concentration of gaseous and vapour-phase organic substances in emissions from stationary sources.