Analyseur de mesure fixe
SIGNAL S4 PULSAR – NDIR Multi-gas analyser
Caractéristiques
Range: See tables (CO, CO₂, CH₄, HCl, N₂O, NO, SO₂)
Resolution: 1000:1
Linearity: 0.5% of range or 1% of reading
Repeatability: 1% of reading
Zero drift: 1% FS per week
Span drift: 1% FS per 24 hours
Warm-up time: Operational in 15 minutes, 1 hour to achieve
specifications
Flow sensitivity: 1% of range per l/min
Sample filter: 10 microns
Output signal: TCP/IP
RS232/AK protocol
0.10 VDC
Optional: 4-20 mA
CANbus
Water interference (H₂O): <1 ppm/2 ppm H₂O typical
Remote operation: Complete software suite running on
Windows VB
Sample temperature: 5°C – 40°C non-condensing
Power supply: 110/220V 50/60 Hz
24 VDC
70–350 Watts depending on the number of channels
Construction materials: 316 stainless steel / PTFE / Quartz / Calcium
fluoride
Dimensions: 19" x 133.5 mm x 530 mm
Weight: Max 23 kg
Options: • Wireless tablet • USB port, SD card • 4–
20 mA output • CANbus • 35 programmable NO-contacts
(can be configured to select a calibration gas per channel and per
range with a calibration gas value defined in the device’s memory)
Oxygen: • Servo-controlled deadweight detector Range 0–5
%, 0–10 %, 0–25 %
• Accuracy: ±0.01% of reading
• Zero drift: ±0.002% O₂ per hour
• Tilt effect: 0.01% O₂ per degree
• Magnetic effect: A mass of soft magnetic material placed anywhere on the housing
will cause a reading variation of less than ±0.1% O₂
• Detector noise: Less than 0.005% O2
• Linearity: Better than 0.1% O2
Points forts
Accurate:
- Rotary gas filter for minimal cross-sensitivity
- Suitable for low-concentration measurement applications
Flexible and reliable:
- Available in 1, 2, 3, 4 or 5-channel gas configurations
- Multiple gas and measurement range options
Easy to use:
- Software suite for operation via Ethernet or RS232
- Wireless tablet
GASES:
- CO / CO2 / CH4 / NO / N2O / SO2 / HCl / O2
Applications:
- CEMS
- Research
- Compliance
- Gas purity
- Automotive
- Air quality
- Process
- Combustion
A wide range of user-defined alarms is available for conditions such as:
1. Concentration limit (user-defined)
2. Sample flow rate (outside limits)
3. Pump failure
4. Heater failure
5. Voltage out of range
6. Thermocouple failure
7. Configuration error
8. Incorrectly configured options
9. Infrared motor failure
10. Infrared source failure
11. Incorrect calculations (no calibration defined)
CO
Measurement range | Resolution | LDL
________________________________________
CO2
Measurement range | Resolution | LDL
________________________________________
CH4
Measurement range | Resolution | LDL
HCl
Measurement range | Resolution | LDL
________________________________________
N2O
Measurement range | Resolution | LDL
NO
Measurement range | Resolution | LDL
________________________________________
SO2
Measurement range | Resolution | LDL
Fourni avec
Series of NDIR multi-gas analysers for industrial gas applications, combustion research, emissions monitoring and gas purity (with optional oxygen analyser).