Analyseur de mesure fixe
SIGNAL S4 AURORA : Analyseur d’oxygène paramagnétique
Caractéristiques
Technical Specifications P1.1/P1.1E
Detector Type: Paramagnetic
Ranges: Range A: 0–5%, 0–10%, 0–25% or
Range B: 0–100% or
Range C: 95–100%
Noise 0.005% O2
Linearity ± 0.1% O2
Repeatability < ±0.01% O2
Drift None <± 0.2% O2 over 30 days
Response Time < 4.5 s
Flow Effect: 0.05% variation in reading with a flow rate variation of 0.4 l/min to 1.4 l/min at atmospheric pressure.
Temperature Effect on Zero: 0.1% O2 variation per °C
Temperature Effect on Span: 0.05% O2 variation per °C
Tilt Effect 0.01% per 2° of tilt
External Magnetic Effect A mass of soft magnetic material anywhere on the housing will cause a variation of less than 0.1% O2 in the reading
Construction 316 stainless steel and quartz
Dimensions 19” x 133.3 mm x 550 mm Weight: 12 kg
Power requirements 110 VAC – 250 VAC or 24 VDC. 70 Watts.
Outputs 0–10 volts
RS232
Ethernet (remote software included)
4-20 mA option.
Gas calibration services Nitrogen (to set the analyser zero)
Oxygen in nitrogen, concentration suitable for calibration. Or ambient air.
Optional Barometric
Atmospheric
Pressure compensation
NEW - Each S4 analyser can now be supplied with a rugged wireless tablet that connects wirelessly to the analyser via a built-in 802.11 Wi-Fi network with a range of up to 50 metres. This allows users to view live data from a different location, and even manage data logging, alarms and calibration
Points forts
Flexible:
Several different configurations
Programmable contact closure for gas circuit control
Easy to use:
Wireless tablet
Software suite for use via Ethernet or RS232
Accurate:
O2 resolution of 0.01%
0.1% linearity
Fourni avec
Paramagnetic oxygen analysers for precise percentage measurements in combustion applications, inert atmospheres, medicine and human safety, internal combustion engine exhaust gases and many other applications.