SIGNAL S4 AURORA : Analyseur d’oxygène paramagnétique

Veuillez entrer votre adresse email pour télécharger la fiche

×

Caractéristiques

Technical Specifications P1.1/P1.1E

Detector Type: Paramagnetic

Ranges: Range A: 0–5%, 0–10%, 0–25% or

Range B: 0–100% or

Range C: 95–100%

Noise 0.005% O2

Linearity ± 0.1% O2

Repeatability < ±0.01% O2

Drift None <± 0.2% O2 over 30 days

Response Time < 4.5 s

Flow Effect: 0.05% variation in reading with a flow rate variation of 0.4 l/min to 1.4 l/min at atmospheric pressure.

Temperature Effect on Zero: 0.1% O2 variation per °C

Temperature Effect on Span: 0.05% O2 variation per °C

Tilt Effect 0.01% per 2° of tilt

External Magnetic Effect A mass of soft magnetic material anywhere on the housing will cause a variation of less than 0.1% O2 in the reading

Construction 316 stainless steel and quartz

Dimensions 19” x 133.3 mm x 550 mm Weight: 12 kg

Power requirements 110 VAC – 250 VAC or 24 VDC. 70 Watts.

Outputs 0–10 volts

RS232

Ethernet (remote software included)

4-20 mA option.

Gas calibration services Nitrogen (to set the analyser zero)

Oxygen in nitrogen, concentration suitable for calibration. Or ambient air.

Optional Barometric

Atmospheric

Pressure compensation


NEW - Each S4 analyser can now be supplied with a rugged wireless tablet that connects wirelessly to the analyser via a built-in 802.11 Wi-Fi network with a range of up to 50 metres. This allows users to view live data from a different location, and even manage data logging, alarms and calibration

Points forts

Flexible:

Several different configurations

Programmable contact closure for gas circuit control

Easy to use:

Wireless tablet

Software suite for use via Ethernet or RS232

Accurate:

O2 resolution of 0.01%

0.1% linearity

Fourni avec

Paramagnetic oxygen analysers for precise percentage measurements in combustion applications, inert atmospheres, medicine and human safety, internal combustion engine exhaust gases and many other applications.