3010 MINIFID : Analyseur de gaz COV FID chauffé portable

Veuillez entrer votre adresse email pour télécharger la fiche

×

Caractéristiques

Technical specifications P1.1/P1.1E

Certifications:

EN 14181 QAI 1 (application pending)

EN 15267-4:2017

EN 12619:2013

EN 61000-6-1:2007

EN 61000-6-3:2007+A1:2011

EN 61000-6-4:2007+A1:2011

EN 61010-2-081:2015

USEPA: Method 25A of the Federal Register 40 CFR Part 60

China: CPA 2015-C174

Sensitivity 0.1% of range

Linearity 2% of measured value

Drift ± 0.2 ppm or 2% of range per week, whichever is greater

Noise ± 0.1 ppm or 1% of range, whichever is greater

T90 response time within 2 seconds

Power supply 110V AC or 220V AC, 500 Watts

Heated sampling line temperature 190 °C

Analyser sampling system temperature 190 °C

Weight with gas cylinders 12 kg (25 lbs)

Weight without gas cylinder 9 kg (18 lbs)

Dimensions 210 x 300 x 240 mm

Safety lock-out Fuel shut-off in the event of flame failure

Sample Lockout The sampling pump and heated line cannot start until the analyser has reached the set temperature

Sample Filter 0.4 microns (accessible at the rear of the analyser)

Sample flow rate 0.4–3 l/min (internal sampling pump)
Sampling line lengths of 5 or 10 metres of 6 mm Teflon tubing heated to 190 °C and controlled by the analyser. The mains plug is wired so that the 5-metre and 10-metre lines connect to different contacts. Other lengths up to 15 metres available on request

Easy transport Shoulder strap with fastening clips

Operating controls 3½-digit LCD display. Temperature readout for the heated line and the sampling system. Power switch.

Range selection and calibration controls

RECORDER OUTPUTS Isolated 0-10 V DC and 4-20 mA

DATA LOGGER Integrated into the analyser with PC software enabling recording in a spreadsheet

Gas quick-connect fittings Colour-coded for fuel and calibration gas

Gas cylinder capacity 1 litre water capacity

H2 fuel consumption 35 ml/min

H2/He fuel consumption 180 ml/min

Calibration gas flow rate 0.4 l/min

Zero air Internally generated with a carbon or catalytic purifier

Flame air Generated internally using a carbon or catalytic purifier

Analyser weather protection IP 20 (IP 55 when fitted with a custom-made fabric cover)

Operating ambient temperature +5 to +40°C

Points forts

Compliance

USEPA: Method 25A of the Code of Federal Regulations 40 CFR Part 60.

China: CPA 2015-C174.

Applied for EN14181 QAL1 (MCERTS).

Flexible and reliable:

Designed for portability.

Military-style design.

Thousands sold worldwide.

Many units still in operation after years of rigorous use.

Easy to use:

20-minute warm-up time.

No microprocessor or software.

Just 5 simple functions.

0–10 V output or optional recorder.

GASES:

Total hydrocarbons

Methane (with optional cutter)

Non-methane (with optional cutter)

Applications:

Regulatory compliance

Verification and calibration of installed emission measurement systems (EMS) in accordance with EN 14181

Temporary backup when permanent EMS are not operational

Process optimisation

Verification of emission reduction performance

Detection of hydrocarbon gas leaks

The analyser is available with readings displayed in mg/m³ and with certified measurement ranges of 0–15/30 and 500 mg/m³.

3010 MINIFID: Portable heated FID VOC analyser


Options :
1. Power supply: Isolation transformer from 230V AC to 110V AC (compliant with UK IEE standards for on-site use)

2. Hydrogen fuel version.

3. “PURE” catalytic air purifier (for the lowest ranges)

4. Measurement range of 0–10% (lowest range of 0–100 ppm)

5. On-board data logger (with software for downloading to a spreadsheet)

6. Heated line options:

a. 5 metres, 110V, 1/4 inch

b. 5 metres, 220V, 1/4 inch

c. 10 metres, 110V, 1/4 inch

d. 10 metres, 220V, 1/4 inch

7. Gas cylinders (empty):

a. 1-litre gas cylinder (water capacity) for FUEL (empty for transport)

b. 1-litre gas cylinder (water capacity) for CALIBRATION GAS (empty for transport)

8. Cylinder pressure regulator for FUEL

9. Cylinder pressure regulator for CALIBRATION GAS

10. Gas cylinders (filled):

a. 1-litre gas cylinder filled to pressure with a mixture of 40% H2 / 60% He

b. 1-litre gas cylinder filled to pressure with calibration gas

c. 1-litre gas cylinder pressurised with H2 fuel

11. Detachable cylinder cage (for fuel and calibration gas)

12. Stainless steel probe with sintered filter

13. Spare activated carbon scrubber (field-replaceable)

14. Non-methane hydrocarbon cutter

An external non-methane hydrocarbon cutter is available for customers who wish to measure methane only. This also enables the calculation of non-methane values

Fourni avec

Portable flame ionisation detector (FID) analysers.
Typical applications include testing emissions from solvent-based processes, combustion processes, engine emissions and VOC abatement systems, as well as the verification and calibration of installed continuous emissions monitoring systems (CEMS) and the provision of a temporary backup solution when CEMS are out of service.